Temporary Bi- Polar Pacing Catheter Size 110Cm, 10 Cm Depth Marker Hemocompatible Polyurethane Coating Stainless Steel Center Conductor Unshrouded/Exposed Pin Connector Leads Etc.
2
Temporary Bi- polar pacing catheter size 110cm, 10 cm depth marker hemocompatible polyurethane coating stainless steel center conductor unshrouded/exposed pin connector leads. Particulars of specification are as follows: 1. Catheter must be Smooth and must having Hemocompatible polyurethane coated 2. Catheter must have stainless steel center conductor and electrodes 3. Catheter must have 10 cm depth markers clearly marked terminals for enhance placements 4. Connector leads must have unshrouded/exposed pin 5. Catheter must be designed to provide pacing and sensing in ventricle and to balance maneuverability & mechanical stability. ]