Provision Of Additional Air Compressor, Air Dryer, Vacuum Pump, Secondary And Reserve Oxygen Supply System In Oxygen Manifold At Aiims Bibinagar, Telangana Provision Of Additional Air Compressor, Air Dryer, Vacuum Pump, Secondary And Reserve Oxygen Supply , Supply, Installation, Testing And Commissioning Of Oxygen Manifold Supply System (Without Cylinders) All Copper Pipe Should Be Suitable To Withstand Working Pressure Of 145 Kg/Cm2. All Sizes Of Components Shall Be Compatable To 2X10 And 2X5 Duplex Manifold And Existing Medical Gas Pipeline Sytem. Total Work Shall Executed As Per Direction Of Engineering - In-Charge. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Supply And Installation Of High Pressure Copper Pipes Of Size 1/2 Inchnormal Bore And 15 Standard Wire Gauge. Including Clamping, Pipe Fittings And Brazing Work , Supply And Fixing Of Brass Through Hole Manifold Block With Lock Nut To Ms Frame. The Distance Between Block To Block Min 300 Mm.Compatable To 2X10 And 2X5 Duplex Manifolds , Supply And Fixing Of Brass One Side Hole (End) Manifold Block With Lock Nut To Ms Frame. The Distance Between Block To Block Min 300 Mm.Compitable To 2X10 And 2X5 Duplex Manifolds. , Supply And Fixing Of 1/4 Inch High Pressure Ball Valve Type Isolation Non Brazable Valves, Brass Bodied Along With Ball And Isolation Valve Fittings, Stem, Stem Seal, Teflon Seal And ‘O’ Ring. , Supply And Fixing Of High Pressure Non-Return Valves Along With Valve Fittings. Compatable To 2X10 And 2X5 Duplex Manifolds , Supply And Fixing Of High Pressure Copper Tail Pipes Of Size 1/4 Inch Nominal Bore And 15 Standard Wire Gaugewith One End Having (Bull Nose) Brass Adapter Suitable For Oxygen Cylinder And Other Suitable For Manifold Nr Valves (Minimumlength 1 Meter). Including Pipe Fittings And Brazing. , Supply And Fabrication Of Mild Steel 50Mm X 50 Mmx 6 Mm L-Section For Fixing Through Hole Manifold Block, One Side Hole (End) Manifold Block Andsupporting Manifold Sysytem. Includingdrilling, Welding, Painting With 2 Coats Of Metal Oxide Primer, Suitable Enamel Paint And Fastening Nut And Bolts. , Supply And Fabrication Of Mild Steel 75Mm X 40 Mmx 6.8 Mm C-Section For Supporting Manifoldsytem. Includingdrilling, Welding, Painting With 2 Coats Of Metal Oxide Primer, Suitable Enamel Paint And Fastening Nut And Bolts. , Supply And Fixing Wall Anchor Bolt-Angar Fastener Rust Proof Rawal Bolt For Concrete And Stone Walls (M10) 10Mm. , Supply And Fabrication Of Galvanized Iron Chain Heavy Duty Made In 6 Mm Rod Chain For Holding The Cylinder,Including Welding, S-Hook, Fastening Nut And Bolts. , Supply And Installation Of 12Mm Copper Pipe For Connection In Between New Manifold/Control Panel Setup To Existing Pipe Line System With All Fittings. , Supply And Installation Of 24 Mm Copper Pipe For Connection In Between New Manifold/Control Pannel Setup To Existing Pipe Line System With All Fittings. Including Brazing And Enamel Paint As Per Colour Code. , Supply And Installation Of 12 Mm Copper Pipe For Connection In Between New Manifold To Control Pannel Setup. , Supply, Installation, Testing And Commissioning Of Medical Gas Alarm For Line Pressure (Min 3 Services O2, Medical Air And Vacuume) To Be Connect Toexisting Pipe Line System : Pressure Monitoring Distribution System Shall Be Done Through Digitally. Alarm Box Shall Be Made Using M.S Duly Powder Coated. System Shall Be Digitally Indicate Pressure. The Digital Alarm Shall Be Safe For Low-Voltage Operation, And The System Shall Be Supported By An Audiovisual Alarm Facility. The Digital Sensor Shall Be Mounted Locally. Pressures Shall Be Indicated In Psi, Kpa Or Bar Read-Out. System Shall Be High, Low Alarm Set Point Are Adjustable For Each Gas Service. System Shall Be True Digital Led Read-Out With Red “Alarm”, Normal Green Trend Indicator For Each Service. Repeat Alarm Shall Beadjustable 1 To 60 Minutes. System Shall Be Up To Minimum 03 Gases In Standard Configuration. Alarm Buzzer Sound Shall Be In Excess Of 90 Decibels. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Supply, Installation, Testing And Commissioning Of Single Phase 220/240 V Medical Gas Fully Automatic Control Panel: Maximum Pressure 200 Bar, Output Pressure 3-7.5 Bar (Adjustable), Inlet Connections Left Bank And Right Bank. The Automated Control Panel System Shall Contain An Analog Pressure Meter To Display The Pressure Of Each Incoming Bank And Mainline Gas Pressure. As Soon As The Gas In One Bank Runs Below The Set Pressure, It Has To Sound An Alarm And Switch To The Other Bank. The System Shall Be Indicate The Specific Colored Led Lights To Indicate The Bank Is In Use And On Standby. The Device Also Needs To Be Indicates Whether The Gas Pressure Is In Normal Or Low Along The Banks And The Mains Outlet Connection, Safety Valve Opening Pressure Primary Pressure 15 Bar, Secondary Pressure 7.5 Bar, Controlpannel Includes Isolation Valves, Line Relief Valve, Manual Purge Valve, Intermediate Relief Valve,Right Line Regulator, Left Line Regulator,Check Valves, Right Bank Dome Regulator, Pressure Transducers, Solenoid Valve, Intermediate Gauge, Left Bank Regulator, Bank Gauge, Line Gauges And All Neccessary Requirements. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Supply, Installation, Testing And Commissioning Of Compressor Setup With Motor Set For Medical Air With Following Specifications: Compressor Technology: Reciprocating Compressor. Medical Air Shall Be Drive By Directly Fan Cooled Electric Motor 15Hpwith “V” Belt Driven. Belt Driven System With Driver And Driven Pulleys Along With Belt Guard. Low Level Protection, Low Maintenance Single Stage Design. Inclding Stater And Cables From Source To Compressor Setup. Compressed Air To Be Connected To Existing Tank With Proper Pipes/Flexible Pipes Andfittings. Type :Base Mounted Oil Free Reciprocating Compressor Without Tank. Motor Type: 15 Hp/ 11 Kw, 3-Phase, Foot Mounted. Free Air Delivery : 1200-1500 Lpm., Air Quality:Class-0Oil Free Air, Working Pressure : 7-10 Kg/Cm2, Minimum Air Discharge Capacity : 40 Cfm, Cooling Type : Air Cooled, No Of Cylinders: 2 Or Above, No Of Stages: 2, Maximum Motor Speed:1410 To 1440 Rpm, Voltage / Ph / Frequency : 415V / 3 Ph / 50 Hz, Minimum Compressor Speed : 600-1000, Starter Type: Automatic Star Delta, Pressure Switch On : 6.5 Kg/Cm2 (Adjustable), Pressure Switch Cut-Off : 8.5 Kg/Cm2 (Adjustable) Type Of Suction: Filter Dry Type (Polyester), Ball Valve, Non Return Valve,Make:Ingersoll Rand/Anest Iwata/Atlascopco, Siemens Energy, Warranty:2 Year Relocation Of Existing Compressor Setup, Including Fitting And Accessories Required For Relocation Of Compressor Setup. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Supply, Installation, Testing And Commissioning Of Of 10 Hpmedical Vacuum Pump With Motor Set Following Specifications: Medical Vacuum Pump Should Be Air Cooled, Oil Lubricated Base Mount Reciprocatingsuitable For Both Continuous And Frequently Start/Stop Operation At Inlet Vacuum Reservoir Between 525Mmhg To 700Mmhg Pressure Strictly As Per Htm 02- 01/En/Din/Nfpa99c. Medical Vacuum Pump Characteristic Must Be Providing For Various Inlet Pressures And Fully Tested And Comply With Htm 2022/Bs/En/Iso7396-1. Pump Mechanism Should Be Smooth And Quiet With Noise Level Below 7 Db At 1 Meter Distance. Medical Vacuum Pump Shall Be Drive By Directly Fan Cooled Electric Motor 10Hpwith “V” Belt Driven. Belt Driven System With Driver And Driven Pulleys Along With Belt Guard. Low Level Protection, Low Maintenance Single Stage Design. Medical Vacuum Pump Must Be Suitable For Central Supply Pipeline System.Medical Vacuum Pump Shall Be Provided Continuous Supply Of Specified Volumetric Flow Rate Of Medical Vacuum To A Pipeline System In Health Care Facilities. Incldingrequired Capacitystater And Cables From Source To Compressor Setup. To Be Connected To Existing Tank With Proper Pipes/Flexible Pipes Andfittings Provision Of 10 Meters Mild Steel 2 Inch Diameter, Medium Class Pipe For Exhaust From Vaccuum Pump (New Setup And Old Vacuum Setup) To Outside Of Manifold Room Including Nonreturn Valves Ball Valves Fittings Bends,Couplings, Welding. Relocation Of Existing Vacuim Pump Setup, Including Fitting And Accessories Required For Relocation Ofvacuim Pump Setup Setup. , Specifications: Make/ Model:Ingersoll Rand / Anest Iwata / Frank Air Type:Reciprocating Type Oil Lubricated. Number Of Piston:2/3 Nos. Maximum Vacuum Capacity:(Inchhg) 29.02”Hg / 741Mmhg Uses/Application:Medical Purpose In Hospital Voltage 440 Volt Automation Grade:Fully Automatic Finishing:Polished / Smooth Maximum Flow Rate:130 Cfm To 148 Cfm Powers:10Hp Above Phases: 3 Phase Warranty: 2 Years Cooling System: Air Cooled. Frequency: 50 Hz/60Hz Maximum Motor Speed:1410 To 1440 Rpm Voltage / Ph / Frequency : 415V / 3 Ph / 50 Hz Starter Type:Automatic Star Delta Auto Pressure Switch On : Adjustable Auto Pressure Switch Cut-Off :Adjustable Tee Joint Fitting During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Provided. Fright And Handling Charges Are Included In The Scope Of The Contractor Only. , Bacteria Filter: Bacteria Filter Plant Shall Fitted With Duplex Bacteria Filter System. Each Individual Filter Shall Have Capacity To Deliver Full Design Flow Such That One Set Is Designated Duty And The Other Will Be Standby. Bacteria Filter Shall Have Efficiency At Least 99.99% When Tested In Accordance With Bs 3928 During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Pre Filter With Automatic Float Drain Outlet Filter Element:Borosilicate Construction Material:Extruded Aluminum Alloy Coating - External: Epoxy Powder Coating Particle Removal: 5 (?M) Max. Oil Carryover:5 (Mg/M3). Max. Working Temp.80 Degree Celsius Initial Pressure Loss:0.03 (Kg/Cm2) Pressure Drop For Element Change :0.4 (Kg/Cm2) Make: Trident/ Ingersoll Rand/ Atlas Copco During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Air Dryer : Supply Installation Testing And Comissioning Of 60 Cfm Heatless Air Dryer : 60 Cfm Heatless Air Dryer : It Should Have Air Dryer Desiccant Type With1” End Bsp Connection. It Should Be Designed For Iso:7183- 1986 (E) Standards. It Should Have Dryer Quality Class -Iso 8573-1 Standards. It Should Be Made Of Aluminum Construction. It Should Have Purge Loss 15+/- 1%. Dew Point -30Pdp Above Make: Trident/ Ingersoll Rand/ Atlas Copco. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Line Filter - 1: Line Filter Shall Automatic Drain Supply, Installation, Testing And Commissioning Of Coarse Coalescing Filter 1 Of 50Cfm To 60 Scfm Filter :It Should Have Flow-Optimized Design For Advanced Filter Head Design For Optimized Flow Performance. It Should Have Flexible Installation Modular Design And Accessible Fixings Enable Simple Close Coupling Assembly. It Should Comply With Air Quality Standard Iso 8573-1: 2010. It Should Have Profiled Bowl Design And Push Fit Elements Ensure Quick And For Reliable Maintenance. It Should Have Corrosion Protection Internal And External Electrophoretic Paint Finish Followed By A Tough Exterior Polyester Powder Coating. It Should Have Color Coded Element End Caps Of Red Color For Easy And Accurate Grade Identification.It Should Have Element Of E30612 Grade. It Should Have Particle Retention Of Up To 99.999%, And Significantly Reduced Pressure Loss. It Should Have Pipe Size Connection 1” = 60Scfm And 90Nm3/Hr. Filter Element:Borosilicate Construction Material:Extruded Aluminum Alloy Coating - External: Epoxy Powder Coating Particle Removal: 1 (?M) Max. Oil Carryover:0.5 (Mg/M3). Max. Working Temp.80 Degree Celsius Initial Pressure Loss:0.15 (Kg/Cm2) Pressure Drop For Element Change :0.4 (Kg/Cm2) Make: Trident/ Ingersoll Rand/ Atlas Copco During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Line Filter - 2: Line Filter Shall Automatic Drain Of Fine Coalescing Filter 1 Of 50Cfm To 60 Scfm Filter :It Should Have Flow- Optimized Design For Advanced Filter Head Design For Optimized Flow Performance. It Should Have Flexible Installation Modular Design And Accessible Fixings Enable Simple Close Coupling Assembly. It Should Comply With Air Quality Standard Iso 8573-1: 2010. It Should Have Profiled Bowl Design And Push Fit Elements Ensure Quick And For Reliable Maintenance. It Should Have Corrosion Protection Internal And External Electrophoretic Paint Finish Followed By A Tough Exterior Polyester Powder Coating. It Should Have Color Coded Element End Caps Of Blue Color For Easy And Accurate Grade Identification.It Should Have Element Of E30612 Grade. It Should Have Particle Retention Of Up To 99.999%,And Significantly Reduced Pressure Loss. Filter Element:Activated Carbon Construction Material:Extruded Aluminum Alloy Coating - External: Epoxy Powder Coating Particle Removal: 0.01 (?M) Max. Oil Carryover:0.01 (Mg/M3). Max. Working Temp.80 Degree Celsius Initial Pressure Loss:0.15 (Kg/Cm2) Pressure Drop For Element Change :0.4 (Kg/Cm2) Make: Trident/ Ingersoll Rand/ Atlas Copco During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Electrical Requirements For Air Compressor, Vacuum Pump, Air Dryer And Filters. During The Execution Of Contract Work. Any Item Which Is Technically Required Not Mentioned In Boq, Need To Be Provided By The Contractor. No Extra Charges Will Be Paid For Such Items. , Supplying And Making End Termination Ofwith Brass Compression Gland And Aluminium Lugs For 4 X 25 Sq. Mm (28Mm) Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required. , Supplying And Fixing 100 Arating, Four Pole, 415 Volts, Isolator In The Existing Mcb Db Complete With Connections, Testing And Commissioning Etc. As Required. Make:Schneider Electric /L&T/Abb/Havells , Providing And Fixing 125 A 16 Ka Rating And Breaking Capacity Mccb Withthermo Magnetic Release And Terminal Spreaders In Existing Cubicle Panel Board Including Drilling Holes In Cubicle Panel, Making Connections, Etc. As Required. Make:Schneider Electric /L&T/Abb/Havells , Supplying And Fixing 6 Way, Double Door , Single Pole And Neutral, Sheet Steel, Mcb Distribution Board, 240 Volts, On Surface/ Recess, Complete With Tinned Copper Bus Bar, Neutral Bus Bar, Earth Bar, Din Bar, Interconnections, Powder Painted Including Earthing Etc. As Required. (But Without Mcb/Rccb/Isolator) Make:Schneider Electric /L&T/Abb/Havells , Supplying And Laying Ofpvc Insulated And Pvc Sheathed / Xlpe Power Cable Of 1.1 Kv Grade Size Of 4 Core X 25 Sq Mm Armoured Aluminium Conductor Cable Make:Polycab /Havells/Finolex/Kei.