Supply, Installation and commissioning Semi automatic Oil break down voltage tester, Capacity 0 -100KV, Technical Specification 1. Input 230 Volts AC-50/60 Hz Voltage. 2. Output 100 KV AC, 3. Resolution 0.1KV, Precise and accurate in measurement of oil BDV as per IS 6792-1992, Test to IS 6792-1992 IEC 156,Manual motorized operation, LCD display, Backlit, Removable mushroom electrode as per IS 6792-1992, Castresin high voltage transformer designed for Min 100 K test cycles, Smooth motorized raise of voltage at 2KV persecond, HT hood interlock for user protection, Breakdown of test object/Liquid at the instance of first spark, Fuse,Use of standard components all confirming to CE standards, Internal test cum calibration report- calibration onHV probe, Operation manual, Warranty Certificate of unit, Go/No-Go gauge 2.5mm. [ Warranty Period: 30Months after the date of delivery ]