Specifications For High Perfomance Liquid Chromatography (Hplc) For The Department Of Clinical Pharmacology, Byl Nair Hospital, Mumbai.-, Mumbai-Maharashtra

Municipal Copr Of Greater Mumbai has published Specifications For High Perfomance Liquid Chromatography (Hplc) For The Department Of Clinical Pharmacology, Byl Nair Hospital, Mumbai.-. Submission Date for this Tender is 08-12-2023. Autosamplers Tenders in Mumbai Maharashtra. Bidders can get complete Tender details and download the document.




Tender Notice

40914617
Specifications For High Perfomance Liquid Chromatography (Hplc) For The Department Of Clinical Pharmacology, Byl Nair Hospital, Mumbai.-
Open Tender
Indian
Maharashtra
Mumbai
08-12-2023

Tender Details

Specifications For High Perfomance Liquid Chromatography (Hplc) For The Department Of Clinical Pharmacology, Byl Nair Hospital, Mumbai.-1. Pump a Operating Principle Parallel- type/ or series type double plunger b Gradient formation Quaternary Low-pressure gradient c Flow rate range 0.001-10 ml/min d Flow rate accuracy ±0.1 - ±1.0% e Flow rate precision ≤0.05 - 0.075%RSD f Pressure range 5000 –10,000psi g No. of eluent lines 4 solutions h Solvent degassing Built – in, minimum 4 channels i Safety Measures Liquid leakage sensor j Concentration accuracy 0.50-0.80%RSD k Composition precision ≤0.25% 2. Autosampler injector with cooler a Sample capacity Minimum 100 × 1.5 ml b Sample injection system Loop injection or total volume sample injection/ direct injection c Sample injection volume 0.1 – 100 μl d Injection volume accuracy ±1.0 % e Injection volume precision ≤0.5 % RSD (when 5 is injected) f Carry over of sample ≤ 0.005 – 0.05% with Caffeine/ Chlorhexidine g Sample thermostatting 4 to 400C h Safety Measures Liquid leakage sensor i Injection cycle time ≤20 seconds j No. of samples for processing 100 or more 3. Column Oven a Type of column oven Forced-air-circulation type / Peltier type b Temperature range 10–85℃ or more c Temperature Accuracy ± 0.1℃ d Temperature precision ≤ 0.2℃ e Column capacity Minimum 2 columns f Safety Measures Liquid leakage sensor 4. Fluorescence detector a Data acquisition ≥70 HZ b Light Source Xenon lamp, minimum life of 2000 working hours c Wavelength Range Excitation -200 -650nm Emission -250-700 nm d Spectral Bandwidth 20nm e Wavelength Accuracy ±3nm f Safety Measures Liquid leakage sensor 5. Photo Diode Array (PDA) detector a Provision of Photo diode array Yes b PDA detector Wavelength range in nm 190 – 800 or better c No. of diodes 1024 d PDA detector light source Deuterium and Tungsten lamps e PDA detector wavelength accuracy ±1nm f Data rate in Hz 100 g Noise in AU Less than 0.8 x 10-5 h PDA detector Drift in AU/hour Less than 0.5 x 10-3 i Temperature controlled flow cell in 0C 50C above room temp or 500C 6. Column a Type of column C-18 b No. of columns 4 c Dimensions of column 250mm x 4.6mm x 5μ

Key Value

Document Fees
Refer document
EMD
Refer document
Tender Value
Refer document