Supply Of Sterilization Indicators, Labels And Catheters In Maximum Of 3 Lots On As And When Required Basis On Pro-Rata Basis Mode , Chemical Indicators Steam, Class – 6 For Steam Sterilization Process Monitoring At121 And 134 Degree Celcius Temperature Fixed Pre Vacuum Cycles. Should Be As Per En-Iso 11140-1:2005, Class/Type 6 Standards (Should Pass Test As Per En Iso 11140-4:2007 For Steamindicators), Colour Change Reference Should Be Printed On Strip. Stated Values, Parameters Should Be Printed On Strip. Manufacturer Details, Class/Type Of Indicator And Iso Number Should Be Clearly Printed Preferably On The Individual Indicator Strip Or The Carton And Should Be Submitted/Mentioned In Offer, Else Will Be Rejected. Shelf Life Shall Be Minimum 02 Years From Date Of Supply. Manufacturer/ Authorized Supplier Only To Quote. 04 Samples To Be Submitted To Aso, Hospital Zonal Stores, Basement, Barc Hospital, Anushaktinagar, Barc Colony, Mumbai 400094 For Analysis In Due Date. , Biological Indicator (Bi) For Steam Sterilization. Self Contained, Bi For Monitoring Of Steam Sterilization, 10^6 Population Of Batcerial Sopres Of Geo Bacillus Stearothermophilus, High D Value Will Be Preffered In The Standards Range, Self Containing Plastic Vials/Ampules For Steam Sterilization At 121 Degree Celcius And 134 Degree Celcius Sterilization Cycles. With Class 5/6 Steam Chemical Indicator Strictly As Per Iso-11140 Standards. All Valid Analysis & Standards Certificates To Be Provided, Should Pass Test As Per En Iso 11140-4:2007 For Steam Indicators.For En 285 Compliant Steam Sterilizers. Manufacturer Name, Exp Date, Standards, Time, Steam/Concentration, Temp. Ratio, Class Of Indicator, Colour Change,Population, All Details Should Be Clearly Printed On Indicator Or On Outer Pack And All Details Should Be Sumbitted/Mentioned In Offer, Else Will Be Rejected. 04 Samples To Be Submitted To Aso, Hospital Zonal Stores, Basement, Barc Hospital, Anushaktinagar, Barc Colony, Mumbai 400094 For Analysis, In Due Date.Manufacturer/ Authorized Supplier Only To Quote. , 3-Line Lable With Class 1 Steam Process Indicator, To Be Used In Pre Vacuum Steam Sterilizers At 134/121 Degree Celsius, For Documentation. To Be Used In Mechanical Dispensing Labeling Gun. Indicator Should Be Lead Free. Approximate Size: 3Cm X 2.5Cm Three Line Double Label With Adhesive, Alpha-Numeric And Numeric Letters Imprinted With Class -1 Chemical Indicator Manditorily, As Per En Iso 11140-1 Standards. Supplied In Rolls, Approximately 600-800 Labels Per Roll. Returnable Labelling Gun With Ink Cartridges Shall Be Supplied With Labels For Use. Shelf Life Shall Be Minimum 2 Years From Date Of Supply. Labels Quantity Per Roll Should Be Mentioned. Manufacturer/ Authorized Supplier Only To Quote. 10 Labels To Be Submitted As Samples In Due Date , Plain - Red Rubbersuction Catheters, International Quality, Sulphur Free, For Medical Use. Made Of Non Toxic, Non Allergic, Treated Rubber, Plain, Perfect Circular Shape Only, Uniform Thickness,No Deformed Shaped Lumen & Catheter Will Be Allowed . Length – 40 Cms Needed. The Closed End Should Besmooth Circular Having Opening At 1Cm. Uniform Length For All Sizes. Manufacturer/ Authorized Supplier Only To Quote.03 Catheters To Be Submitted As Samples In Due Date. Required Size Quantities: No. 3- 600, No.4 – 600, No.5 – 1300Nos, No.6 - 1700Nos, No.7 - 3100Nos, No.8 - 3100Nos, No.9 - 3000Nos, No.10 - 1000Nos, No.11- 300, No. 12- 300 Total Quantity: 15000 Nos