| 1 | HIGH FREQUENCY GENERATOR Generator should be of latest technology with high frequency 50KHz X-Raygenerator. Constant Power output of 30 KW.KV range should be 40 to 125 KV in 1KV/step.mA output: 400mAmAs range should be 1 to 200mAs |
| 2 | X-RAY TUBE:i. A Dual focus Rotating anode X-ray tube should be provided.ii. Focal spot size of 1.0mm (small focus) x 2.00mm (large focus).iii) Anode heat storage capacity should be 140kHU or moreiv) Collimator having halogen lamp / bright light source and auto shut provision of the light.v) HV Cable: 1 Pair of H.V. Cable of suitable length. |
| 3 | STAND:Ceiling Free tube Stand with Counter Balanced Tube Head with following features should be provided. Tube Stand: ? Height of the stand: 2200mm or Less ? Longitudinal movement of column on track: 1950mm or more ? Total up/down movement of the tube head: 1200mm or more.Detector Stand: ? Height of the stand: 2200mm or Less ? Up /down travel of Bucky: 1200mm or more ? Bucky tilting angle: 0° to 90° to be provided |
| 4 | TABLE:Mobile Diagnostic table should be provided. ? Table should be of following dimensions:- Length: 1900mm or more- Width: 660mm or more- Height: 721mm or less ? Locks should be available on front wheels for table stability during exposure. ? Maximum weight carrying capacity for the table should be 180Kg or more. |
| 5 | FLAT PANEL DETECTOR: Specifications:The detector should be flat panel type with A-Si (amorphous silicon) and CSi as scintillator.Size of detector must be 43cm x 43cm.Active Image matrix 3K x 3K.Image depth should be 14bit.Pixel size should be 150um or less (Smaller pixel size is preferred)Detector resolution should be more than 3.3 lp/mm.DQE (Detector Quantum Efficiency) should be more than 65%. |
| 6 | X-RAY/IMAGECONTROL CONSOLEFully integrated system with following features:- Digital Display of KV & mAs.- KV & mAs increase and decrease control on G.U.I (Graphicaluser interface).- Ready and X-Ray ON indication on G.U.I (Graphicaluser interface).- Self diagnostic Program for error code display of faults such asEarth fault error, KV error, Filament error & Tube’sThermal Overload.- An Inbuilt overload protection device.- Anatomical Programming Radiography (i.e. APR): Preprogrammed parameters of human Anatomy which helps the user to select exposure parameters based on body part, examination view and size of the patient. Since it is a computer based system (full system integration) so any number of Organ programming combinations is possible. User can define his own Organ parameters and can edit the existing parameters to his satisfaction and comfort level.- APR programs: More than 1000 programs. (Expandable asper user’s requirement).A dual action hand Switch with Retractable cord for RadiationProtection of Operator. |
| 7 | IMAGE ACQUISITION SOFTWARE AND ITS CHARACTERISTICS:Software provides complete control of all image capture functions within the examination room. It enhances the entire workflow by delivering diagnostic images instantly. It also allows user to transfer X-Ray images electronically to remote workstations, image archives, and printers, also has an excellent performance on image quality control such as: M A I N F E A T U R E S • D i g i t a l i m a g e p r o c e s s i n gt e c h n o l o g y ? Preview image in less than 5 seconds. ? Exam Specific Algorithms image processing forconsistent image quality of all body parts. ? Preset image processing tools for different anatomy ? Image cropping ? Image mirror, rotate. ? Image annotation ? Add image accept/reject comments ? Rejected images archival with provision of convertingthem to Accepted images. ? Separate log for Rejected, Accepted and Printed images. ? True size for printing. ? Hard disc capacity for image storage > 3000 images. ? Inbuilt CD/DVD writer facility ? Remote online system diagnosis ? Remote online software upgrade ? Image quality control tools ? Easy and quick Offset and gain calibration with badpixel removal algorithm. |
| 8 | MONITOR: 1 No. 19” LCD TFT monitor should be provided. |
| 9 | OTHER IMPORTANT REQUIRMENTS:- The company should be ISO and ICMED certified.- The unit should be approved by AERB.- The company should have proven track record in Govt. sector. – - The Detector, Generator & Imaging software should be from same principal manufacturer of the X-ray system. |
| 10 | Printer SpecificationParameters SpecificationsPrint Technology Direct thermal (dry, daylight-load film)Spatial Resolution 320dpi (12.6 pixels/mm)Processing Capacity ? 14\"x17\" > 70 sheets/h? 8\"x10\" > 90 sheets/hGreyscale Contrast Resolution 14 bitsFilm Tray 2 Trays with 100 Films capacity each.Film Sizes 8\"x10\", 10\"x12\", 11\"x14\", 14\"x17\"Applicable film Medical Dry Thermal Film (blue or clear base)Interface 10/100/1000 Base-T Ethernet (RJ-45)Network Protocol Standard DICOM 3.0 connectionImage Quality Automatic calibration using built-in densitometerControl Panel Touch Screen, Online Display, Alert, Fault and ActivePower Supply 100-240VAC 50/60Hz 600WWeight 50KgDimensions 740 x 566 x 445 mmOperating Temperature 5°C to 35°COperating Humidity 35%-85%Storage Humidity 30%-95%Storage Temperature -22°Cto 50°CSoftware ? VNC Printer Configuration Software? Minimum System Configuration Requirement? Windows 10 or higher? 4GB RAM or higher? 500 HDD or higher |
| 11 | Power Backup UPS 3KVA On-Line UPS |
| 12 | Air Conditioner 1.5 ton Split AC, 5 Star and Stabilizer |