| 1 | Oxygen manifold shall be of size 5+5 bulk cylinders. Manifold shall consist of two high-pressure header bar assemblies to facilitate connection of primary and secondary cylinder supplies. Each header bar shall be provided with 10 numbers of cylinder pigtail connections to suit cylinder valves as per standards incorporating a check valve at the header connection. The high-pressure header bar shall be designed in such a manner that it can be extended to facilitate additional cylinder connections. Each header bar assembly shall be provided with a high-pressure shut-off valve.The cylinder should be placed with the help of cylinder brackets and fixing chains which should be zinc plated. Manifold including Copper Tail Pipe, NRV, bull nose fittings and cylinder bracket. |
| 2 | Oxygen Emergency manifold with regulator shall be of size 4+4 bulk cylinders. Manifold shall consist of two high-pressure header bar assemblies to facilitate connection of primary and secondary cylinder supplies. Each header bar shall be provided with 8 numbers of cylinder pigtail connections to suit cylinder valves as per standards incorporating a check valve at the header connection. The high-pressure header bar shall be designed in such a manner that it can be extended to facilitate additional cylinder connections. Each header bar assembly shall be provided with a high-pressure shut-off valve.The cylinder should be placed with the help of cylinder brackets and fixing chains which should be zinc plated. Manifold including Copper Tail Pipe, NRV, bull nose fittings and cylinder bracket. |
| 3 | Semi Automatic Control Panel for the Oxygen Manifold system complete for automatic changeover from one bank to other bank & vice versa, with all necessary hardware accessories etc. It should fully complies and meets with the requirements of Health Technical Memorandum 02-01 (HTM 02-01) standards. |
| 4 | Two (1+1) Cylinder Emergency Oxygen Manifold with a high flow regulator with gauges and safety valves complete with all accessories etc.Emergency with Middle Frames Two Cylinder Oxygen Manifold should be suitable to withstand a pressure of 145Kg/cm2, along with high-pressure copper annealed tail pipes with end Brass adapters suitable for Oxygen Cylinders and manifold. A High Pressure Regulator to be mounted on the Manifold System for reducing the cylinder pressure suitable to the line pressure. |
| 5 | Gas specific Terminal (Oxygen) outlets with probes/adapters complete as per specifications.All terminal units should conform to BS 5682:1984 (1992). Terminal units intended for installation with the socket axis horizontal, ie wall mounted, include a non-swivel device so that directly connected equipment such as flow meters remain vertical; terminal units intended for installation with the socket axis vertical, for example in certain types of pendant, do not include a non-swivel device. |
| 6 | Medical graded Copper pipes( Maxflow/Rajco/Mandev or equivalent) shall be solid drawn, tempered, seamless, phosphorous deoxidized, non-arsenic and degreased for oxygen service. Distribution Copper Pipe manufactured as per BSEN:13348:2001. Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. All exposed pipes should be painted with two coats of synthetic enamel paint and colour codification. 28mm OD x 0.9 mm thk |
| 7 | Medical graded Copper pipes( Maxflow/Rajco/Mandev or equivalent) shall be solid drawn, tempered, seamless, phosphorous deoxidized, non-arsenic and degreased for oxygen service. Distribution Copper Pipe manufactured as per BSEN:13348:2001. Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. All exposed pipes should be painted with two coats of synthetic enamel paint and colour codification. 22mm OD x 0.9 mm thk |
| 8 | Medical graded Copper pipes( Maxflow/Rajco/Mandev or equivalent) shall be solid drawn, tempered, seamless, phosphorous deoxidized, non-arsenic and degreased for oxygen service. Distribution Copper Pipe manufactured as per BSEN:13348:2001. Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. All exposed pipes should be painted with two coats of synthetic enamel paint and colour codification. 15mm OD x 0.9 mm thk |
| 9 | Medical graded Copper pipes( Maxflow/Rajco/Mandev or equivalent) shall be solid drawn, tempered, seamless, phosphorous deoxidized, non-arsenic and degreased for oxygen service. Distribution Copper Pipe manufactured as per BSEN:13348:2001. Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. All exposed pipes should be painted with two coats of synthetic enamel paint and colour codification. 12mm OD x 0.7 mm thk |
| 10 | The Medical Vacuum System Single pump shall ensure the minimum pipeline. Vacuum level of 450mmHg is maintained at the plant service connection point at the rated volumetric. The bacteria filtration system shall be fitted in pipeline. Vacuum pumps shall be air-cooled, oil lubricated rotary vane type suitable forcontinuous and frequent start/stop operation at nominal inlet vacuum levels of between 570mmHg and 720mmHg approx. Central Vacuum system with Single Vacuum pump (IR/Aneast-Iwata) of vacuumof 745mm, with three phase Motor 5.0 HP, storage tank of 1000 Litre, and all required accessories vacuum gauge, vacuum pressure switch, Silencer etc. Cascading of vacuum pumps shall be achieved by measuring the vacuum level at the plant inlet with a pressure transducer. A mechanical back-up facility shall ensure continued operation in the event of a control system malfunction. |
| 11 | Vacuum Filter station imported ( Domnick Hunter/ Nitrogenium/Trident Make or equivalent ). These filters are optimised for high-efficient removal of bacterial and other contamination (solids and liquids) from the suction side of vacuum pumps preventingdamagetothe pumpandthepotentialbiologicalinfectionofthe surroundingenvironment.Removedliquidsare collectedinatransparentflask which can be removed for sterilisation.Theefficiencyoftheinstalledfilterelementsexceedsthe0,005%penetration specifiedinHTM2022forinfectiousdiseaseunits,whentestedinaccordancewith BS 3928. |
| 12 | Gas specific Terminal (Vacuum) outlets with probes/adapters are designed & manufacturerd as per BS EN13485.Outlets shall be manufactured with a Copper inlet pipe stub which is silver brazed to the outlet body. Body shall be of one piece brass construction. For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall have break safe mechanism and also comply to EN 737 pressure test standards and rated at minimum 200 psi in the event the primary check valve is removed for maintenance. The outlet assembly should have separate colour coding for each services and will accept onlycorresponding gas specific adaptors. All outlets shall be cleaned and de-greased for medical gas service, factory assembled and tested. The medical gas outlets should be of quick connecting and wall mounted modular type. |
| 13 | Central Air system with Single air compressor oil free 5.0 HP with motor make ABB/CG complete with 3 stage filtration system, Heatless dryer system, air regulating system with air tank capacity of 1000 Litre. |
| 14 | Compressed Air Terminal Outlets with probes/adapters complete as per specifications.Terminal (Air-4 Bar) outlets with probes/adapters are designed & manufacturerd as per BS EN13485. |
| 15 | Creative 3 Gas Alarm Panel to indicate any abnormality of gas pressures and other failures of the system. The alarm system shall be complete with display, sensor module and power supply. The alarm shall be complete with all indications controls, wirings, accessories etc. complete as required.3 Gas Alarm Panel with Analog gauge Display & HI-LOW-NORMAL status. |
| 16 | Area Valve Box - 3 Gas Services with complete fittings, shaddles, Pipes, Gauge Analog, without Valves. Valve box housing must be made of galvanized steel duly powder coated. |
| 17 | ISOLATION VALVE complete. The isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service.All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply.28 mm Valve with Brass Nut & Nipple |
| 18 | ISOLATION VALVE complete. The isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service.All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply.22 mm Valve with Brass Nut & Nipple |
| 19 | ISOLATION VALVE complete. The isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service.All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply.15 mm Valve with Brass Nut & Nipple |
| 20 | ISOLATION VALVE complete. The isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service.All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply.Butterfly Valve with Brass Nut & Nipple |
| 21 | Oxygen BPC Flowmeter with Humidifier (0-15 litres/minute) with adapter, tubing etc. Back Pressure Compensated flow meter will be of accurate gas flow measurement with followingfeatures:Control within a range of 0 – 15 lpm (calibration within +/- 10% ).It will meet strict precision and durability standard.The flow tube and shroud components shall be made of clear, impact resistant polycarbonate.Flow Tube shall have large and expanded 0 – 5 lpm range for improved readability at low flows. The humidifier bottle shall be made of unbreakable polycarbonate material and autoclavable at 121 degree Centigrade temperature. |
| 22 | Ward Vacuum Unit 600mL wall mounted type. Ward Vacuum Unit shall be of light weight and compact. The unit will consist of a regulator, 600 ml Reusable collection jar, made ofunbreakable poly carbonate material and fully autoclavable at 134 degree centigrade.A wall bracket for mounting the jar assembly on the wall.The vacuum regulator shall be infinitely adjustable and have vacuum gauge which indicates suction supplied by the regulator. Safety trap shall be provided inside the jar to safeguard the regulator from overflowing. |