Purchase Of Consumable Items For Cardiology Department On Two Years Rate Contract Basis ( Group D ) . - 2 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 1: Intravenous Cannula For Neonates N• Available In 24 And 26 G Sizes N• Individually Packed With Easy Peel Off N• Needle Should Have Protective Plastic Tube Covering N 3 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 2: Intravenous Cannula For Children & Adults N• Available In 16, 18, 20, 22 G N• 20 G Should Allow The Passage Of 0.025-Inch Guide Wire N• Individually Packed With Easy Peel Off N• Needle Should Have Protective Plastic Tube Covering N 4 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 3: Puncture Needle For Neonates & Children N• 20- 22G N• 3- 5 Cm Long N• Should Allow Passage Of 0.021 Or 0.025-Inch Guide Wire N• Should Be Supplied Individually Packed N• Needle Should Have Protective Plastic Tube Covering N 5 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 4: Micropuncture Needle With Stainless Steel Guidewire For Neonates & Children N• 20- 22 G N• 3- 5 Cm Long N• Should Include 0.018-Inch Stainless Steel Introducer Guide Wire N• Should Be Supplied Individually Packed N• Needle Should Have Protective Plastic Tube Covering N 6 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 5: Micropuncture Needle With Hydrophilic Guidewire For Neonates & Children N• 20- 22 G N• 3- 5 Cm Long N• Should Include 0.018-Inch Hydrophilic Guide Wire N• Should Be Supplied Individually Packed N• Needle Should Have Protected Plastic Tube Covering N 7 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 6: Micropuncture Introducer Set For Neonates & Children N• 20-22 G N• 3-5 Cm Long Needle N• Should Include 0.018-Inch Guide Wire N• Should Include 4- 5 Fr Coaxial Sheath N• Should Include A 3 Fr Dilator N• All Components Should Be Individually Packed N• Needle Should Have Protective Plastic Tube Covering N 8 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 7: Introducer Sheath For Neonates N• Must Be Available In 3 Fr Size N• Should Include 0.14 Or 0.18-Inch Guidewire N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 9 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 8: Introducer Sheath With Introducer Wire For Children N• 4-6 Fr Sizes N• 5.5-7.5 Cm Long N• 0.021 Or 0.025-Inch Introducer Guide Wire N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 10 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 9: Introducer Sheath With Straight Introducer Wire For Children N• 4-6 Fr Sizes N• 5.5-7.5 Cm Long N• 0.021 Or 0.025-Inch Straight Tip Introducer Guide Wire N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 11 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 10: Introducer Sheath With J Tip Introducer Wire For Children N• 4- 6 Fr Sizes N• 5.5- 7.5 Cm Long N• 0.021 Or 0.025-Inch Introducer Guide Wire With J Tip On One End And Straight Tip On The Other End With Both Ends Suitable For Insertion In The Vessel N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 12 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 11: Introducer Sheath With Hydrophilic Guidewire For Children N• 4-6 Fr Sizes N• 5.0-7.5 Cm Long N• 0.021 Or 0.025-Inch Hydrophilic Introducer Guide Wire With Straight Or J Tip N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 13 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 12: Special Introducer Sheath With Extrathin Wall N• Should Have A Thin Wall So That Od Of The Sheath For 5 Fr Is = 2.14 Mm, 6 Fr = 2.46 Mm & 7 Fr = 2.79Mm N• With Hemostatic Valve To Prevent Back Leak Of Blood And Aspiration Of Air N• With Integrated Side Arm With Attached 3-Way Stopcock N• With Suture Eye For Securing Sheath N• Kink Resistant N• With Dilator-Hub Lock Mechanism To Prevent Its Back-Out During Insertion N• Should Have Smooth And Resistance Free Insertion N 14 Group D1- Puncture Needle & Introducer Sheath For Neonates & Children Nitem No. 13: Chiba Or Chiba Type Puncture Needle For Fetal Interventions N• 16- 22 G Sizes N• 15-20 Cm Long N• Should Have Beveled Tip To Provide Steerability N• Should Have Good Visibility Under Ultrasound N• Should Be Supplied Individually Packed N• Needle Should Have Protective Plastic Tube Covering N 15 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 1: Special Diagnostic Catheters For Neonates N• Preferably Available In Sizes Starting From 3 Fr N• Available In Different Shapes, Curves And Lengths Including But Not Limited To Jr, Jl, Mpa And Pigtail Catheter N• Good Visualization Under Fluoroscopy N• Kink Resistant 16 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 2: Pigtail Catheter For Children N• 4-5 Fr Sizes N• With Multiple Side Holes ( 6 To 12 ) Located Spirally At The Distal End For Uniform Contrast Delivery N• For Safety Reasons, The Loop Of The Pigtail Must Be Retained When Using In Vivo N• Good Visualization Under Fluoroscopy N• Kink Resistant N 17 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 3: Angled Pigtail Catheter For Children N• 4-5 Fr Sizes N• Angle Between Shaft And Distal End Of 145 To 155° N• For Safety Reasons, The Loop Of The Pigtail Must Be Retained When Using In Vivo N• Good Visualization Under Fluoroscopy N• Kink Resistant N 18 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 4: Marker Pigtail Catheter For Children N• 4-5 Fr Sizes N• Must Have Equidistant Multiple Radio Opaque Markers N• For Safety Reasons, The Loop Of The Pigtail Must Be Retained When Using In Vivo N• Good Visualization Under Fluoroscopy N• Kink Resistant N 19 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 5: Judkins Catheter For Children N• 4-5 Fr Sizes N• Left And Right Judkins Catheters In Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 20 Group D2 Ndiagnostic Catheters For Neonates & Children Nitem No. 6: Special Judkins Coronary Catheter With Smaller Curves For Children N• 4-5 Fr Sizes N• Must Be Available In 1.5, 2.0, 2.5 And 3.0 Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 21 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 1: Multipurpose Catheter For Children N• 4- 5 Fr Sizes N• Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 22 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 2: Picard Catheter For Children N• 4- 5 Fr Sizes N• Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 23 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 3: Cobra Catheter For Children N• 4- 5 Fr Sizes N• Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 24 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 4: Cournard Catheter For Children N• 4- 6 Fr Sizes N• Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 25 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 5: Nih Catheter For Children N• 4- 7 Fr Sizes N• Various Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 26 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 6: Balloon Floation Catheter ( Swan Ganz Type ) For Neonates And Infants N• Should Be Available In Various Standard Sizes N• The Catheter Should Have A Low Profile For Optimal Usage In Neonates And Infants N• 4 Fr Must Be Compatible With 4 Fr Introducer Sheath N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 27 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 7: Balloon Floation Catheter ( Swan Ganz Type ) For Children N• Should Be Available In Various Standard Sizes N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 28 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 8: Balloon Tipped Angiography ( Berman ) Catheter For Neonates And Infants N• Should Be Available In Various Standard Sizes N• The Catheter Should Have A Low Profile For Optimal Usage In Neonates And Infants N• 4Fr Must Be Compatible With 4 Fr Introducer Sheath N• Should Have 6 - 8 Holes Proximal To The Balloon For Dye Injection N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 29 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 9: Balloon Tipped Angiography ( Berman ) Catheter For Children N• Should Be Available In Various Standard Sizes N• Should Have 6 - 8 Holes Proximal To The Balloon For Dye Injection N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 30 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 10: Reverse Berman Catheter N• Should Be Available In Various Standard Sizes N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Should Have Holes Distal To The Balloon For Dye Injection N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 31 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 11: Special Multitrack Type Catheter For Simultaneous Angiography And Pressure Measurement N• 3- 6 Fr Shaft Sizes N• Should Be Compatible With 5-8 Fr Introducer Sheath N• Should Have A Short Distal Extension With A Lumen For Guide Wire N• Should Permit Simultaneous Angiography And Pressure Measurement Keeping The Guidewire In Situ N• Should Be Compatible With 0.025 To 0.035 Inch Guidewire N• Good Visualization Under Fluoroscopy N• Kink Resistant N 32 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 12: Diagnostic Catheters With Special Hydrophilic Coating N• Should Be Available In 4 Fr And 5 Fr Sizes N• Should Have Special Hydrophilic Coating ( M Polymer Coat ) For Variable Lengths At The Distal End Of Catheter For Better Trackability And Navigation N• Should Have Single Or Double Braided Stainless-Steel Construction N• Should Be Compatible With 0.021 To 0.038 Inch Guidewire N• Should Be Available In All Standard Curves And Lengths N• Good Visualization Under Fluoroscopy N• Kink Resistant N 33 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 13: Renal Double Curve Catheter N• 4- 5 Fr Sizes N• Available In Various Standard Lengths And Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 34 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 14: Head Hunter Catheter N• 4- 5 Fr Sizes N• Available In Various Standard Lengths And Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 35 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 15: Simmons / Sidewinder Catheter N• 4- 5 Fr Sizes N• Available In Various Standard Lengths And Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 36 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 16: Sos Omni Selective Catheter N• 4- 5 Fr Sizes N• Available In Various Standard Lengths And Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 37 Group D3 Nspecial Diagnostic Catheters For Neonates & Children Nitem No. 17: Coeliac Axis Catheter N• 4- 5 Fr Sizes N• Available In Various Standard Lengths And Curves N• Good Visualization Under Fluoroscopy N• Kink Resistant N 38 Group D4 Nguiding Catheters For Neonates & Children Nitem No. 1: Judkins Right Coronary Guiding Catheter For Neonates & Children N• Should Be Available In 5 Fr N• Must Be Available In Various Standard Lengths And Curves N 39 Group D4 Nguiding Catheters For Neonates & Children Nitem No. 2: Renal Guiding Catheter For Neonates & Children N• Should Be Available In 5 Fr N• Must Be Available In Various Standard Lengths N 40 Group D4 Nguiding Catheters For Neonates & Children Nitem No. 3: Multipurpose Guiding Catheter For Children Nshould Be Available In 5 Fr N• Must Be Available In Various Standard Lengths And Curves N 41 Group D4 Nguiding Catheters For Neonates & Children Nitem No. 4: Special Guiding Catheters ( Other Than Coronary Curves ) N• Should Be Braided With Low Friction Inner Layer N• Should Have Atraumatic Soft Tip N• Should Be Available In Various Standard Tip Shapes And Curves For Access To Different Parts Of The Peripheral Vasculature, Including Other Standard Shapes N• Should Be Available In Od Of 5-8 Fr N• Should Have The Largest Id For Each Od N• Should Be Available In All Standard Lengths N 42 Group D4 Nguiding Catheters For Neonates & Children Nitem No. 5: Guiding Catheters With Hydrophilic Coating For Neonates And Children N• Should Be Braided With Low Friction Inner Layer N• Should Have Atraumatic Soft Tip N• Should Be Available In All Standard Tip Shapes And Curves For Access To Different Parts Of The Peripheral Vasculature N• Should Be Available In Od Of 5-8 Fr N• Should Have The Largest Id For Each Od N• Should Be Available In All Standard Lengths N• Should Have A Hydrophilic Coating To Provide Least Resistance During Insertion And Withdrawal N 43 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 1: Mullins Sheath Without Side Arm Of 4- 9 Fr Sizes For Structural Intervention N• Should Have Thick And Stiff Dilator For Guiding Septal Puncture 44 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 2: Mullins Sheath With Side Arm: 4- 8 Fr N• Must Have A Check Valve With Side Arm N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 45 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 3: Mullins Sheath With Side Arm: 4- 8 Fr N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With = 0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 46 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 4: Mullins Sheath With Side Arm: 9- 12 Fr N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With = 0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 47 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 5: Mullins Sheath With Side Arm: > 12 Fr Size N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With = 0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 48 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 6: Long Sheath With Side Arm: Hausdorf Curve N• Must Be Available In Various Standard Sizes N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With = 0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 49 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 7: Long Sheath With Side Arm: Balkin Curve N• Must Be Available In Various Standard Sizes N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With = 0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 50 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 8: Special Long Sheaths With Side Arm: Other Than Mullins Curve N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With 0.018” Wire, 0.025” Wire Or 0.035” Wire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 51 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 9: Special Braided Long Sheaths With Side Arm: Other Than Mullins Curve N• Should Have An Additional Braided Layer For Better Flexibility And Kink Resistance N• Must Have Multiple Dilators Compatible With 0.018” Wire, 0.025” Wire Or 0.035” Wire N• Must Have A Check Valve With Side Arm N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 52 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 10: Long Sheaths With Deflectable Tip: 6 -9 Fr N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With =0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 53 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 11: Long Sheaths With Deflectable Tip: 10-12 Fr N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With =0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 54 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 12: Long Sheaths With Deflectable Tip: >12 Fr N• Must Have A Check Valve With Side Arm N• Dilator Must Be Compatible With =0.035 Inch Guidewire N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 55 Group D5 Nlong Sheath For Valvular And Structural Intervention Nitem No. 13: Long Sheath For Carotid Intervention N• Sheath Without Dilator Must Have Good Radio Opacity For Better Fluoroscopic Visibility N• Must Have A Check Valve With Side Arm N• Should Be Kink Resistant With A Reinforcement Mechanism N• Should Have A Minimum Length Of 90 Cm N• Must Be Available In Various Standard Lengths And Curves N• Colour Coded For Size Identification N• Smooth Transition From Dilator To Sheath N 56 Group D6 Nvalvuloplasty Balloon Nitem No. 1: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Manufactured In India N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 – 0.018-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 57 Group D6 Nvalvuloplasty Balloon Nitem No. 2: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Usfda Approved N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 – 0.018-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 58 Group D6 Nvalvuloplasty Balloon Nitem No. 3: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Ce Marked N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 – 0.018-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 59 Group D6 Nvalvuloplasty Balloon Nitem No. 4: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Manufactured In India N• Must Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 60 Group D6 Nvalvuloplasty Balloon Nitem No. 5: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Usfda Approved N• Must Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 61 Group D6 Nvalvuloplasty Balloon Nitem No. 6: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Ce Marked N• Must Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 62 Group D6 Nvalvuloplasty Balloon Nitem No. 7: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Manufactured In India N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 63 Group D6 Nvalvuloplasty Balloon Nitem No. 8: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Usfda Approved N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 64 Group D6 Nvalvuloplasty Balloon Nitem No. 9: Percutaneous Transluminal Valvuloplasty Balloon ( Low Burst Pressure ) – Ce Marked N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures < 10 Atm N 65 Group D6 Nvalvuloplasty Balloon Nitem No. 10: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 3-4 Fr N• Must Be Compatible With 0.014 – 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 66 Group D6 Nvalvuloplasty Balloon Nitem No. 11: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 3- 4 Fr N• Must Be Compatible With 0.014 – 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 67 Group D6 Nvalvuloplasty Balloon Nitem No. 12: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Ce Marked N• Must Be Compatible With 3- 4 Fr N• Must Be Compatible With 0.014 – 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 68 Group D6 Nvalvuloplasty Balloon Nitem No. 13: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 5- 7 Fr N• Must Be Compatible With 0.025 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 69 Group D6 Nvalvuloplasty Balloon Nitem No. 14: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 5- 7 Fr N• Must Be Compatible With 0.025 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 70 Group D6 Nvalvuloplasty Balloon Nitem No. 15: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Ce Marked N• Must Be Compatible With 5- 7 Fr N• Must Be Compatible With 0.025 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 71 Group D6 Nvalvuloplasty Balloon Nitem No. 16: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 72 Group D6 Nvalvuloplasty Balloon Nitem No. 17: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 73 Group D6 Nvalvuloplasty Balloon Nitem No.18: Percutaneous Transluminal Valvuloplasty Balloon ( Medium To High Burst Pressure ) – Ce Marked N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.035 – 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures 10 - 15 Atm N 74 Group D6 Nvalvuloplasty Balloon Nitem No. 19: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 3- 4 Fr N• Must Be Compatible With 0.014– 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 75 Group D6 Nvalvuloplasty Balloon Nitem No. 20: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 3- 4 Fr N• Must Be Compatible With 0.014– 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 76 Group D6 Nvalvuloplasty Balloon Nitem No. 21: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Ce Marked N• Must Be Compatible With 3-4Fr N• Must Be Compatible With 0.014– 0.025-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 77 Group D6 Nvalvuloplasty Balloon Nitem No. 22: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 5-7 Fr N• Must Be Compatible With 0.025– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 78 Group D6 Nvalvuloplasty Balloon Nitem No. 23: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 5-7 Fr N• Must Be Compatible With 0.025– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 79 Group D6 Nvalvuloplasty Balloon Nitem No. 24: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Ce Marked N• Must Be Compatible With 5-7 Fr N• Must Be Compatible With 0.025– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 80 Group D6 Nvalvuloplasty Balloon Nitem No. 25: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Manufactured In India N• Must Be Compatible With 8 Fr And Higher N• Must Be Compatible With 0.035– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 81 Group D6 Nvalvuloplasty Balloon Nitem No. 26: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Usfda Approved N• Must Be Compatible With 8 Fr And Higher N• Must Be Compatible With 0.035– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 82 Group D6 Nvalvuloplasty Balloon Nitem No. 27: Percutaneous Transluminal Valvuloplasty Balloon ( Very High Burst Pressure ) – Ce Marked N• Must Be Compatible With 8 Fr And Higher N• Must Be Compatible With 0.035– 0.038-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Rated Burst Pressures Greater Than 15 Atm N 83 Group D6 Nvalvuloplasty Balloon Nitem No. 28: Special Percutaneous Transluminal Valvuloplasty Balloon Having Rapid Exchange System N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 – 0.018-Inch Wire N• Monorail / Rapid Exchange Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 84 Group D6 Nvalvuloplasty Balloon Nitem No. 29: Special Percutaneous Transluminal Valvuloplasty Balloon Having Rapid Exchange System Nmust Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Monorail Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 85 Group D6 Nvalvuloplasty Balloon Nitem No. 30: Special Percutaneous Transluminal Valvuloplasty Balloon Having Rapid Exchange System N• Must Be Compatible With 8 Fr Or Higher Sheath N• Must Be Compatible With 0.025 – 0.038-Inch Wire N• Monorail Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 86 Group D7 Nperipheral Angioplasty Balloon Nitem No. 1: Peripheral Angioplasty Balloon Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 Or 0.018-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 87 Group D7 Nperipheral Angioplasty Balloon Nitem No. 2: Peripehral Angioplasty Balloon Compatible With 5-7 Fr Sheath N• Must Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths 88 Group D7 Nperipheral Angioplasty Balloon Nitem No. 3: Peripheral Angioplasty Balloon Compatible With > 7 Fr Sheath N• Must Be Compatible With 8 Fr And Higher N• Must Be Compatible With 0.035-Inch Wire N• Over The Wire ( Otw ) Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 89 Group D7 Nperipheral Angioplasty Balloon Nitem No. 4: Peripheral Angioplasty Balloon Compatible With 3- 4 Fr Sheath Having Rapid Exchange System N• Must Be Compatible With 3- 4 Fr Sheath N• Must Be Compatible With 0.014 Or 0.018-Inch Wire N• Monorail Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 90 Group D7 Nperipheral Angioplasty Balloon Nitem No. 5: Peripheral Angioplasty Balloon Compatible With 5-7 Fr Sheath Having Rapid Exchange System N• Must Be Compatible With 5- 7 Fr Sheath N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Monorail Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 91 Group D7 Nperipheral Angioplasty Balloon Nitem No. 6: Peripheral Angioplasty Balloon Compatible With > 7 Fr Sheath Having Rapid Exchange System N• Must Be Compatible With 8 Fr And Higher N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Monorail Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 92 Group D7 Nperipheral Angioplasty Balloon Nitem No. 7: Special High Burst Pressure Angioplasty Balloon With Hydrophilic Coating N• Should Have Rated Burst Pressure = 14 Atmosphere N• Should Have Special Hydrophilic Coating For Better Trackability N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N 93 Group D8 Nvascular Stents Nitem No. 1: Premounted Balloon Expandable Vascular Stents - Manufactured In India N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N 94 Group D8 Nvascular Stents Nitem No. 2: Premounted Balloon Expandable Vascular Stents – Usfda Approved N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N 95 Group D8 Nvascular Stents Nitem No. 3: Premounted Balloon Expandable Vascular Stents – Ce Marked N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N 96 Group D8 Nvascular Stents Nitem No. 4: Premounted Balloon Expandable Vascular Stents With Rapid Exchange Systems - Manufactured In India N• Approved For Pediatric And Adult Use N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Monorail Positioning System N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N 97 Group D8 Nvascular Stents Nitem No. 5: Premounted Balloon Expandable Vascular Stents With Rapid Exchange Systems – Usfda Approved N• Approved For Pediatric And Adult Use N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Monorail Positioning System N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile 98 Group D8 Nvascular Stents Nitem No. 6: Premounted Balloon Expandable Vascular Stents With Rapid Exchange Systems – Ce Marked N• Approved For Pediatric And Adult Use N• Must Be Compatible With 0.018-Inch Or Lesser Diameter Guidewire N• Monorail Positioning System N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile 99 Group D8 Nvascular Stents Nitem No. 7: Premounted Balloon Expandable Vascular Stents - Manufactured In India N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Compatible With 0.035 Or 0.038-Inch Wire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 100 Group D8 Nvascular Stents Nitem No. 8: Premounted Balloon Expandable Vascular Stents – Usfda Approved N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Compatible With 0.035 Or 0.038-Inch Wire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 101 Group D8 Nvascular Stents Nitem No. 9: Premounted Balloon Expandable Vascular Stents – Ce Marked N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries N• Compatible With 0.035 To 0.038-Inch Wire N• Should Be Over The Wire N• Must Be Available In Various Standard Lengths And Diameters N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 102 Group D8 Nvascular Stents Nitem No. 10: Unmounted Balloon Expandable Vascular Stents- Manufactured In India N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries Etc N• Must Be Available In Various Standard Lengths And Diameters N• Should Be Made Up Of Stainless Steel Alloy Or Cobalt Chromium N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 103 Group D8 Nvascular Stents Nitem No. 11: Unmounted Balloon Expandable Vascular Stents – Usfda Approved N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries Etc N• Must Be Available In Various Standard Lengths And Diameters N• Should Be Made Up Of Stainless Steel Alloy Or Cobalt Chromium N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 104 Group D8 Nvascular Stents Nitem No. 12: Unmounted Balloon Expandable Vascular Stents – Ce Marked N• Approved For Pediatric And Adult Use In Non-Coronary Locations Such As Peripheral Vessels, Pulmonary Arteries Etc N• Must Be Available In Various Standard Lengths And Diameters N• Should Be Made Up Of Stainless Steel Alloy Or Cobalt Chromium N• Should Have Good Flexibility, Deliverability And Low Profile N• Preferably Should Have Less Foreshortening And Ability To Expand Stent At A Later Time N 105 Group D8 Nvascular Stents Nitem No. 13: Premounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Manufactured In India N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Be Premounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time N 106 Group D8 Nvascular Stents Nitem No. 14: Premounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Usfda Approved N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Be Premounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time 107 Group D8 Nvascular Stents Nitem No. 15: Premounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Ce Marked N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Be Premounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time N 108 Group D8 Nvascular Stents Nitem No. 16: Unmounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Manufactured In India N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Not Be Mounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time N 109 Group D8 Nvascular Stents Nitem No. 17: Unmounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Nusfda Approved N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Not Be Mounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time N 110 Group D8 Nvascular Stents Nitem No. 18: Unmounted Balloon Expandable Vascular Stents For Coarctation Of Aorta – Nce Marked N• Should Be Approved For Coarctation Of Aorta ( Coa ) Stenting In Children And Adults N• Must Be Available In Various Standard Lengths And Diameters N• The Stent Must Not Be Mounted On Balloon N• Should Have Good Flexibility, Deliverability And Low Profile N• Must Have Less Foreshortening And Ability To Expand Stent At A Later Time N 111 Group D8 Nvascular Stents Nitem No. 19: Covered Vascular Stents Premounted On Balloon: 5-10 Mm Diameter – Manufactured In India N• Should Be Available In Diameters Of 5, 6, 7, 8, 9 &10Mm N• Should Be Compatible With 6 – 7 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 112 Group D8 Nvascular Stents Nitem No. 20: Covered Vascular Stents Premounted On Balloon: 5-10 Mm Diameter – Usfda Approved N• Should Be Available In Diameters Of 5, 6, 7, 8, 9 &10Mm N• Should Be Compatible With 6 – 7 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 113 Group D8 Nvascular Stents Nitem No. 21: Covered Vascular Stents Premounted On Balloon: 5-10 Mm Diameter – Ce Marked N• Should Be Available In Diameters Of 5, 6, 7, 8, 9 &10Mm N• Should Be Compatible With 6 – 7 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 114 Group D8 Nvascular Stents Nitem No. 22: Covered Vascular Stents Premounted On Balloon: 11- 16 Mm Diameter – Manufactured In India N• Should Be Available In Diameters Of 12, 14 & 16Mm N• Should Be Compatible With 9- 11 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 115 Group D8 Nvascular Stents Nitem No. 23: Covered Vascular Stents Premounted On Balloon: 11- 16 Mm Diameter – Usfda Approved N• Should Be Available In Diameters Of 12, 14 & 16Mm N• Should Be Compatible With 9- 11 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 116 Group D8 Nvascular Stents Nitem No. 24: Covered Vascular Stents Premounted On Balloon: 11- 16 Mm Diameter – Ce Marked N• Should Be Available In Diameters Of 12, 14 & 16Mm N• Should Be Compatible With 9- 11 Fr Sheath N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 117 Group D8 Nvascular Stents Nitem No. 25: Covered Vascular Stents Premounted On Balloon: > 16 Mm Diameter – Manufactured In India N• Should Be Available In Diameter Of 16Mm Or More N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 118 Group D8 Nvascular Stents Nitem No. 26: Covered Vascular Stents Premounted On Balloon: > 16 Mm Diameter – Usfda Approved N• Should Be Available In Diameter Of 16Mm Or More N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 119 Group D8 Nvascular Stents Nitem No. 27: Covered Vascular Stents Premounted On Balloon: > 16 Mm Diameter – Ce Marked N• Should Be Available In Diameter Of 16Mm Or More N• The Stent Must Be Premounted On Balloon N• Should Have Low Crossing Profile N 120 Group D8 Nvascular Stents Nitem No. 28: Vascular Stents Premounted On Balloon-In-Balloon ( Bib ) Catheter N• Must Be Available In Various Standard Lengths And Diameters And Configurations N• Should Be Expandable To A Range Of Diameters N• The Stent Must Be Premounted On Balloon In Balloon Catheter N• Should Have Low Crossing Profile N 121 Group D8 Nvascular Stents Nitem No. 29: Covered Vascular Stents Premounted On Balloon-In-Balloon ( Bib ) Catheter N• Must Be Available In Various Standard Lengths And Diameters And Configurations N• Should Be Expandable To A Range Of Diameters N• The Stent Must Be Premounted On Balloon In Balloon Catheter N• Should Have Low Crossing Profile N 122 Group D8 Nvascular Stents Nitem No. 30: Self- Expandable Covered Vascular Stents – Manufactured In India N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Fabric Coating N• Should Be Available In Various Standard Lengths And Diameters N• Should Have A Shaft Length Of 80-140 Cm. N 123 Group D8 Nvascular Stents Nitem No. 31: Self- Expandable Covered Vascular Stents – Usfda Approved N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Fabric Coating Nshould Be Available In Various Standard Lengths And Diameters N• Should Have A Shaft Length Of 80-140 Cm. N 124 Group D8 Nvascular Stents Nitem No. 32: Self- Expandable Covered Vascular Stents – Ce Approved N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Fabric Coating N• Should Be Available In Various Standard Lengths And Diameters N• Should Have A Shaft Length Of 80-140 Cm. N 125 Group D8 Nvascular Stents Nitem No. 33: Self- Expandable Covered Peripheral Vascular Stents – Manufactured In India N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Ptfe Coating On The Inner Lumen N• Should Be Available In Outer Diameters Of 5-13 Mm N• Should Be Available In Stent Lengths Of 2-15 Cm N• Should Be On A Shaft Length Of 75-140 Cm. N 126 Group D8 Nvascular Stents Nitem No. 34: Self- Expandable Covered Peripheral Vascular Stents –Usfda Approved N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Ptfe Coating On The Inner Lumen N• Should Be Available In Outer Diameters Of 5-13 Mm N• Should Be Available In Stent Lengths Of 2-15 Cm N• Should Be On A Shaft Length Of 75-140 Cm. N 127 Group D8 Nvascular Stents Nitem No. 35: Self- Expandable Covered Peripheral Vascular Stents – Ce Marked N• Should Be 0.035” Compatible N• Should Be Over The Wire N• Should Be Made Of Nitinol With Ptfe Coating On The Inner Lumen N• Should Be Available In Outer Diameters Of 5-13 Mm N• Should Be Available In Stent Lengths Of 2-15 Cm N• Should Be On A Shaft Length Of 75-140 Cm N 128 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 1: Fogarty Embolectomy Catheter For Use In Neonates N• Must Be Available In Various Standard Sizes Including Less Than 4 Fr N• Should Have A Low Profile For Optimal Usage In Neonates And Infants N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Must Have A Radiopaque Straighter Inside For Positioning Of The Balloon N 129 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 2: Fogarty Embolectomy Catheter N• Must Be Available In Various Standard Sizes 4-6 Fr N• Should Have A Low Profile For Optimal Usage In Neonates And Infants N• Catheter Should Be Tapered At Tip To Ensure Uniform Diameter Of The Whole Catheter N• Must Have A Radiopaque Straighter Inside For Positioning Of The Balloon N 130 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 3: Atrial Septostomy Balloon N• Must Be Available In Various Standard Sizes 3-7 Fr N• Should Have A Low Profile And Noncompliant Balloon N• Double Lumen Shaft For Over The Wire Positioning N• Central Lumen Should Accommodate 0.014 – 0.025 Guide Wire N• Should Have A Low Profile For Optimal Usage In Neonates And Infants N• Should Be Radiopaque With Good Visibility On Fluoroscopy N 131 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 4: Balloon In Balloon Catheter N• The Diameter Of Inner Balloon Should Be Half The Outer Balloon Diameter N• The Length Of Inner Balloon Should Be 1 Cm Shorter Than The Outer Balloon Length N• The Balloon Should Be Non-Compliant Made Of Thermoplastic Elastomer N• The Diameter Of Outer Balloon Of Bib Catheter Should Be Available In Sizes 12-30 Mm And Balloon Lengths Of 2.5-6.0 Cm N 132 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 5: Cutting Balloon For Peripheral Angioplasty N• Must Be Available In Various Standard Sizes ( Diameter ) And Lengths N• Must Be Compatible With 0.018 – 0.035-Inch Wire N• Over The Wire Positioning System N• Low Profile N• Allows Rapid Inflation & Deflation N 133 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 6: Inflation Device With Maximum Inflation Pressure Only Upto 15 Atm N• Luminescent Analog Pressure Gauge With Maximum Pressure Up To 15 Atm N• Syringe Volume Should Range Between 15 Ml To 60 Ml N• Should Be Appropriate For Use In Valvuloplasty Procedures N• Efficient Locking System To Maintain High Pressure N• Rapid Inflation And Deflation N• Clear Barrel For Easy Visualization Of De-Bubbling N• Ergonomic And User Friendly Hand Held Design N 134 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 7: Inflation Device Maximum Inflation Pressure Greater Than 15 And Upto 30 Atm N• Luminescent Analog Pressure Gauge With Maximum Pressure Range Greater Than 15 Atm And Upto 30 Atm N• Syringe Volume Should Range Between 15 Ml To 60 Ml N• Should Be Appropriate For Use In Valvuloplasty Procedures N• Efficient Locking System To Maintain High Pressure N• Rapid Inflation And Deflation N• Clear Barrel For Easy Visualization Of De-Bubbling N• Ergonomic And User Friendly Hand Held Design N 135 Group D9 Nspecial Balloon Catheters And Accessories Nitem No. 8: Inflation Device With Maximum Inflation Pressure Greater Than 30 Atm N• Luminescent Analog Pressure Gauge With Maximum Pressure Range Greater Than 30 Atm N• Syringe Volume Should Range Between 15 Ml To 60 Ml N• Should Be Appropriate For Use In Valvuloplasty Procedures N• Efficient Locking System To Maintain High Pressure N• Rapid Inflation And Deflation N• Clear Barrel For Easy Visualization Of De-Bubbling N• Ergonomic And User Friendly Hand Held Design N 136 Group D10 Nclosure Devices – Asd Nitem No. 1: Asd Closure Device With Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 137 Group D10 Nclosure Devices – Asd Nitem No. 2: Asd Closure Device With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 138 Group D10 Nclosure Devices – Asd Nitem No. 3: Asd Closure Device With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 139 Group D10 Nclosure Devices – Asd Nitem No. 4: Asd Closure Device With Existing Fenestration With Delivery System – Manfactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 140 Group D10 Nclosure Devices – Asd Nitem No. 5: Asd Closure Device With Existing Fenestration With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 141 Group D10 Nclosure Devices – Asd Nitem No. 6: Asd Closure Device With Existing Fenestration With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 142 Group D10 Nclosure Devices – Asd Nitem No. 7: Asd Closure Device Without Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 143 Group D10 Nclosure Devices – Asd Nitem No. 8: Asd Closure Device Without Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 144 Group D10 Nclosure Devices – Asd Nitem No. 9: Asd Closure Device Without Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 145 Group D10 Nclosure Devices – Asd Nitem No. 10: Delivery System For Asd Device – Manufactured In India N• Available In All Approved Sizes N• Approved For Use In Children And Adults N• Should Comprise Of 45° Curve Delivery Sheath With A Check Valve Mechanism N• Sheath Should Be Kink Resistant With Reinforcement With Internal Braided Layer Made Of Stainless Steel N• Should Have A Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 146 Group D10 Nclosure Devices – Asd Nitem No. 11: Delivery System For Asd Device – Usfda Approved N• Available In All Approved Sizes N• Approved For Use In Children And Adults N• Should Comprise Of 45° Curve Delivery Sheath With A Check Valve Mechanism N• Sheath Should Be Kink Resistant With Reinforcement With Internal Braided Layer Made Of Stainless Steel N• Should Have A Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 147 Group D10 Nclosure Devices – Asd Nitem No. 12: Delivery System For Asd Device – Ce Marked N• Available In All Approved Sizes N• Approved For Use In Children And Adults N• Should Comprise Of 45° Curve Delivery Sheath With A Check Valve Mechanism N• Sheath Should Be Kink Resistant With Reinforcement With Internal Braided Layer Made Of Stainless Steel N• Should Have A Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 148 Group D10 Nclosure Devices – Asd Nitem No. 13: Special Asd Closure Device For Very Large Asd N• Must Be Available In Sizes Greater Than 40 Mm N• Must Be Approved For Use In Adults N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 149 Group D10 Nclosure Devices – Asd Nitem No. 14: Special Asd Closure Device For Fenestrated Defects N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 150 Group D10 Nclosure Devices – Asd Nitem No. 15: Special Hubless Asd Closure Devices With Bioptome Type ( Non-Screw ) Delivery Mechanism N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Without Hub On The Left Atrial Side Of Device N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Self-Centering Device Compatible With An Angled Delivery System N• Device Connects With Delivery Cable Using A Ball And Socket Connection For Easier And Faster Release N• Must Have An Additional Safety Feature For Recapture Of Device After Release From The Cable N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Pusher Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 151 Group D10 Nclosure Devices – Asd Nitem No. 16: Delivery System For Special Hubless Asd Closure Devices With Bioptome Type ( Non Screw ) Delivery Mechanism N• Available In All Approved Sizes N• Approved For Use In Children And Adults N• Should Comprise Of 45° Curve Delivery Sheath With A Check Valve Mechanism N• Sheath Should Be Kink Resistant With Reinforcement With Internal Braided Layer Made Of Stainless Steel N• Should Have A Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise Should Be Included N• Should Include The Pusher Cables Of All Available Sizes That Is Compatible With Various Standard Sizes Of Devices With Bioptome Type Engagement Mechanism N• The Proximal End Of Pusher Cable Should Be Provided With A Lock And Release Safety Mechanism N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Pusher Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 152 Group D10 Nclosure Devices – Asd Nitem No. 17: Device For Maintening The Patency Of Atrial Septal Defect ( Atrial Flow Regulator ) With Delivery System– Manufactured In India N• Available In Different Fenestration And Waist Sizes N• Approved For Pediatric And Adult Use N• Device Must Be Made Up Of Biologically Inert Material N• Self-Centering, Detachable Double Disk Device With Delivery Cable N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 153 Group D10 Nclosure Devices – Asd Nitem No. 18: Device For Maintenance Of Asd With Delivery System ( Atrial Flow Regulator ) – Usfda Approved N• Available In Different Fenestration And Waist Sizes N• Approved For Pediatric And Adult Use N• Device Must Be Made Up Of Biologically Inert Material N• Self-Centering, Detachable Double Disk Device With Delivery Cable N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 154 Group D10 Nclosure Devices – Asd Nitem No. 19: Device For Maintenance Of Asd With Delivery System ( Atrial Flow Regulator ) – Ce Marked N• Available In Different Fenestration And Waist Sizes N• Approved For Pediatric And Adult Use N• Device Must Be Made Up Of Biologically Inert Material N• Self-Centering, Detachable Double Disk Device With Delivery Cable N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 155 Group D11 Nclosure Devices – Vsd Nitem No. 1: Muscular Vsd Closure Devices With Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 156 Group D11 Nclosure Devices – Vsd Nitem No. 2: Muscular Vsd Closure Devices With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 157 Group D11 Nclosure Devices – Vsd Nitem No. 3: Muscular Vsd Closure Devices With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 158 Group D11 Nclosure Devices – Vsd Nitem No. 4: Muscular Vsd Closure Devices Without Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 159 Group D11 Nclosure Devices – Vsd Nitem No. 5: Muscular Vsd Closure Devices Without Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 160 Group D11 Nclosure Devices – Vsd Nitem No. 6: Muscular Vsd Closure Devices Without Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 161 Group D11 Nclosure Devices – Vsd Nitem No. 7: Asymmetric Vsd Closure Devices With Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 162 Group D11 Nclosure Devices – Vsd Nitem No. 8: Asymmetric Vsd Closure Devices With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 163 Group D11 Nclosure Devices – Vsd Nitem No. 9: Asymmetric Vsd Closure Devices With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 164 Group D11 Nclosure Devices – Vsd Nitem No. 10: Asymmetric Vsd Closure Devices Without Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 165 Group D11 Nclosure Devices – Vsd Nitem No. 11: Asymmetric Vsd Closure Devices Without Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 166 Group D11 Nclosure Devices – Vsd Nitem No. 12: Asymmetric Vsd Closure Devices Without Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 167 Group D11 Nclosure Devices – Vsd Nitem No. 13: Special Double Disc Multipurpose Occluder For Vsd With Delivery System N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Device Must Be Made Up Of Biologically Inert Material N• Central Waist And Double-Disc Design For Occluding Various Defects N• Screwable From Both The Discs To Allow Retrograde And Antegrade Delivery Of Device N• Soft Material High Conformability N• Very Low Profile N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 168 Group D12 Nclosure Devices- Pda Nitem No. 1: Pda Closure Device With Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 169 Group D12 Nclosure Devices- Pda Nitem No. 2: Pda Closure Device With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 170 Group D12 Nclosure Devices- Pda Nitem No. 3: Pda Closure Device With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 171 Group D12 Nclosure Devices- Pda Nitem No. 4: Pda Closure Device Without Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 172 Group D12 Nclosure Devices- Pda Nitem No. 5: Pda Closure Device Without Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 173 Group D12 Nclosure Devices- Pda Nitem No. 6: Pda Closure Device Without Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device Must Be Made Up Of Biologically Inert Material N 174 Group D12 Nclosure Devices- Pda Nitem No. 7: Symmetrical Pda Closure Device With Discs On Both Sides With Delivery System N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Device Should Have A Symmetric Design Enabling Both Antegrade And Retrograde Delivery N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N N 175 Group D12 Nclosure Devices- Pda Nitem No. 8: Symmetrical Pda Closure Device With Retention Disc On Both Sides For Use In Preterm Neonates N• Must Be Available In All Approved Sizes N• Must Be Approved For Use In Neonates And Preterm Neonates N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Device Should Have A Symmetric Design Enabling Both Antegrade And Retrograde Delivery N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 176 Group D12 Nclosure Devices- Pda Nitem No. 9: Delivery System For Pda Device – Manufactured In India N• Approved For Use In Infants, Children And Adults N• Available In All Approved Sizes ( Including 4 And 5 Fr ) N• Should Comprise Of 180° Curve Delivery Sheath With Check Valve Mechanism N• Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 177 Group D12 Nclosure Devices- Pda Nitem No. 10: Delivery System For Pda Device – Usfda Approved N• Approved For Use In Infants, Children And Adults N• Available In All Approved Sizes ( Including 4 And 5 Fr ) N• Should Comprise Of 180° Curve Delivery Sheath With Check Valve Mechanism N• Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 178 Group D12 Nclosure Devices- Pda Nitem No. 11: Delivery System For Pda Device – Ce Marked N• Approved For Use In Infants, Children And Adults N• Available In All Approved Sizes ( Including 4 And 5 Fr ) N• Should Comprise Of 180° Curve Delivery Sheath With Check Valve Mechanism N• Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N• Available In All Approved Sizes For Adult And Pediatric Usage N 179 Group D12 Nclosure Devices- Pda Nitem No. 12: Delivery System For Pda Device With Symmetrical Disc On Both Sides N• Approved For Use In Infants, Children And Adults N• Compatible With 4 And 5 Fr Sheath N• Should Comprise Of 180° Curve Delivery Sheath With Check Valve Mechanism N• Soft Radiopaque Tip N• Dilator With Tapering Tip Should Fit In The Sheath Locking Mechanism N• Loader, Plastic Vise And Delivery Cable Should Be Included N 180 Group D12 Nclosure Devices- Pda Nitem No. 13: Special Cylindrical Pda Closure Devices With Delivery System N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Should Have Cylindrical Shape Unlike Amplatzer Type Pda Occluder N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 181 Group D12 Nclosure Devices- Pda Nitem No. 14: Pda Closure Devices Without Hub On The Aortic End Of Device With Delivery System N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Without Hub In The Aortic End Of Device N• Should Have Conical Shape But With Pa End Larger Than Aortic End N• Should Be Available In Different Waist Length Variants N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Pusher Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 182 Group D12 Nclosure Devices- Pda Nitem No. 15: Pda Closure Devices For Large Size Pdas N• Must Be Approved For Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Amplatzer Duct Occlude Type Devices With Aortic End Of Device Measuring Greater Than 24 Mm N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 183 Group D12 Nclosure Devices- Pda Nitem No. 16: Special Double Disc Multipurpose Occluder For Pda N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Central Waist And Double-Disc Design For Occluding Various Defects N• Screwable From Both The Discs To Allow Retrograde And Antegrade Delivery Of Device N• Soft Material High Conformability N• Very Low Profile N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 184 Group D13 Nclosure Devices - Pfo Nitem No. 1: Pfo Closure Devices With Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 185 Group D13 Nclosure Devices - Pfo Nitem No. 2: Pfo Closure Devices With Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 186 Group D13 Nclosure Devices - Pfo Nitem No. 3: Pfo Closure Devices With Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 187 Group D13 Nclosure Devices - Pfo Nitem No. 4: Pfo Closure Devices Without Delivery System – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N 188 Group D13 Nclosure Devices - Pfo Nitem No. 5: Pfo Closure Devices Without Delivery System – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N 189 Group D13 Nclosure Devices - Pfo Nitem No. 6: Pfo Closure Devices Without Delivery System – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N 190 Group D13 Nclosure Devices - Pfo Nitem No. 7: Pfo Closure Devices Without Hub With Delivery System N• Must Be Available In All Approved Sizes N• Must Be Approved For Adult Use N• Self-Centering, Detachable Device With Pusher Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Without Hub In The Left Atrial Side Of Device N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Pusher Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 191 Group D14 Nvascular Plugs & Closure Devices Nitem No. 1: Vascular Plug ( Single Lobe Design ) – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 192 Group D14 Nvascular Plugs & Closure Devices Nitem No. 2: Vascular Plug ( Single Lobe Design ) – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 193 Group D14 Nvascular Plugs & Closure Devices Nitem No. 3: Vascular Plug ( Single Lobe Design ) – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 194 Group D14 Nvascular Plugs & Closure Devices Nitem No. 4: Vascular Plug ( Multisegmented And Oblong Design ) – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 195 Group D14 Nvascular Plugs & Closure Devices Nitem No. 5: Vascular Plug ( Multisegmented And Oblong Design ) – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 196 Group D14 Nvascular Plugs & Closure Devices Nitem No. 6: Vascular Plug ( Multisegmented And Oblong Design ) – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 197 Group D14 Nvascular Plugs & Closure Devices Nitem No. 7: Vascular Plugs For Small Vessels Less Than 15 Mm – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Compatible With 4Fr Guiding Catheter Or 5Fr Diagnostic Catheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration 198 Group D14 Nvascular Plugs & Closure Devices Nitem No. 8: Vascular Plugs For Small Vessels Less Than 15 Mm – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Compatible With 4Fr Guiding Catheter Or 5Fr Diagnostic Catheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration 199 Group D14 Nvascular Plugs & Closure Devices Nitem No. 9: Vascular Plugs For Small Vessels Less Than 15 Mm – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• Compatible With 4Fr Guiding Catheter Or 5Fr Diagnostic Catheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration 200 Group D14 Nvascular Plugs & Closure Devices Nitem No. 10: Vascular Plugs For Very Small Vessels Less Than 5 Mm – Manufactured In India N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Smallest Size Devices Should Be Deliverable By A Microcatheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 201 Group D14 Nvascular Plugs & Closure Devices Nitem No. 11: Vascular Plugs For Very Small Vessels Less Than 5 Mm – Usfda Approved N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Smallest Size Devices Should Be Deliverable By A Microcatheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 202 Group D14 Nvascular Plugs & Closure Devices Nitem No. 12: Vascular Plugs For Very Small Vessels Less Than 5 Mm – Ce Marked N• Must Be Available In All Approved Sizes N• Must Be Approved For Neonatal, Pediatric And Adult Use N• Self-Centering, Detachable Device With Delivery Cable N• Device And All Accessories Must Be Made Up Of Biologically Inert Material N• The Smallest Size Devices Should Be Deliverable By A Microcatheter N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 203 Group D14 Nvascular Plugs & Closure Devices Nitem No. 13: Devices For Paravalvular Leak Closure With Delivery System N• Device Designed And Dedicated For Paravalvular Leak Closure N• Available In All Approved Dimensions And Shapes To Handle Leaks Of Different Morphologies N• Device Must Be Made Up Of Biologically Inert Material N• Repositionable And Retrievable Device N• The Vendor Should Be Able To Supply All The Components Of The Delivery System ( Delivery Sheath And Cable ) N• Must Quote The Price Of Individual Components Of Delivery System ( Delivery Sheath And Cable ) Separately – Attach Pdf N• For Comparative Analysis The Price Of Whole Unit Will Be Taken Into Consideration N 204 Group D15 Naccessories For Device Closure Nitem No. 1: Sizing Balloon For Asd Device Closure: Oval Shaped N• Compliant Balloon Material Flexible Shaft And Soft Tip N• Radio-Opaque Markers Bands N• Available In All Approved Sizes N• Over The Wire: Compatible With 0.035 Wire N 205 Group D15 Naccessories For Device Closure Nitem No. 2: Sizing Balloon For Asd Device Closure: Circular Shape N• Compliant Balloon Material Flexible Shaft And Soft Tip N• Radio-Opaque Markers Bands N• Available In All Approved Sizes N• Over The Wire: Compatible With 0.035 Wire N 206 Group D15 Naccessories For Device Closure Nitem No. 3: Sizing Plate For Balloon Sizing Of Asd N• Measuring Plate With Circular Openings Range From 4 To 38 Mm N 207 Group D16 Nguidewires - Hydrophilic Nitem No. 1: Hydrophilic Guide Wire: Regular Length, Standard Stiffness N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Hydrophilic Coating N 208 Group D16 Nguidewires - Hydrophilic Nitem No. 2: Special Hydrophilic Guide Wire: Regular Length, Standard Stiffness N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 209 Group D16 Nguidewires - Hydrophilic Nitem No. 3: Hydrophilic Diagnostic Guide Wire: Exchange Length, Standard Stiffness N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Hydrophilic Coating N 210 Group D16 Nguidewires - Hydrophilic Nitem No. 4: Special Hydrophilic Guide Wire: Exchange Length, Standard Stiffness N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 211 Group D16 Nguidewires - Hydrophilic Nitem No. 5: Hydrophilic Guide Wire: Regular Length, Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Hydrophilic Coating N 212 Group D16 Nguidewires - Hydrophilic Nitem No. 6: Special Hydrophilic Guide Wire: Regular Length, Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 213 Group D16 Nguidewires - Hydrophilic Nitem No. 7: Hydrophilic Guide Wire: Exchange Length, Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Hydrophilic Coating N 214 Group D16 Nguidewires - Hydrophilic Nitem No. 8: Special Hydrophilic Guide Wire: Exchange Length, Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 215 Group D16 Nguidewires - Hydrophilic Nitem No. 9: Hydrophilic Guide Wire: Regular Length, Extra-Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Short, Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Hydrophilic Coating N 216 Group D16 Nguidewires - Hydrophilic Nitem No. 10: Special Hydrophilic Guide Wire: Regular Length, Extra-Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Short, Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be Between 120-180 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 217 Group D16 Nguidewires - Hydrophilic Nitem No. 11: Hydrophilic Guide Wire: Exchange Length, Extra-Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Hydrophilic Coating N 218 Group D16 Nguidewires - Hydrophilic Nitem No. 12: Special Hydrophilic Guide Wire: Exchange Length, Extra-Stiff N• Must Be Available In 0.018, 0.025, 0.032, 0.035 And 0.038-Inch Size Diameter N• Should Have Super Elastic Alloy Core N• Should Have Super Flexible Wire Tip N• Should Be Available In Straight And Angled Tip N• Should Be 260-300 Cm Long N• Must Have A Special Hydrophilic M Polymer Coating N 219 Group D17 Nother Guidewires Nitem No. 1: Guidewire 0.018 Inch Diameter: Regular Length, Standard Stiffness N• Ptfe Coated Stainless Steel Wire N• Should Have Floppy Tip N• Should Be Available In Straight And J Tip N• Should Be Between 140-180 Cm Long N 220 Group D17 Nother Guidewires Nitem No. 2: Guidewire 0.018 Inch Diameter: Regular Length, Stiff N• Ptfe Coated Stainless Steel Wire N• Should Have Floppy Tip N• Should Be Available In Straight And J Tip N• Should Be Between 140-180 Cm Long N 221 Group D17 Nother Guidewires Nitem No. 3: Guidewire 0.018 Inch Diameter: Exchange Length, Standard Stiffness N• Ptfe Coated Stainless Steel Wire N• Should Have Floppy Tip N• Should Be Available In Straight And J Tip N• Should Be Between 240- 300 Cm Long N 222 Group D17 Nother Guidewires Nitem No. 4: Guidewire 0.018 Inch Diameter: Exchange Length, Stiff N• Ptfe Coated Stainless Steel Wire N• Should Have Floppy Tip N• Should Be Available In Straight And J Tip N• Should Be Between 240-300 Cm Long N 223 Group D17 Nother Guidewires Nitem No. 5: Guidewire 0.035 Inch Diameter: Extra Floppy Exchange Length N• Ptfe Coated Stainless Steel Wire N• Should Have Floppy Tip N• Should Be Available In Straight And J Tip N• Should Be Between 240-300 Cm Long N 224 Group D17 Nother Guidewires Nitem No. 6: Special Guidewire 0.018 Inch Diameter: Extra Floppy Exchange Length N• Ptfe Coated Stainless Steel Wire N• Should Have Extra-Floppy Shaft N• 6 Mm J Tip Noodle Wire Configuration N• 0.035-Inch Outer Diameter Usable Length 260 To 300 Cm N